NeoPICC 1.9 FR Catheter. Percutaneous Implanted long term intravascular catheter. The product is shipped in cases of 5 units. Catalog numbers S1PIC1.9-S, S1PIC1.9-N, S1PIC1.9-SMK, and S1PIC1.9-C.

FDA Recall Z-0349-05 · Posted 2005-01-08

Recall StatusTerminated
Product CodeLJS
Date Initiated2004-11-24
Date Terminated2006-06-14
Product Quantity36340 units

Reason for Recall

complaints of vessel erosion when using this product

Firm Action

The recalling firm issued a letter dated 11/23/04 to physicians informing them of the increase in complaints for vessel erosion and one death (at that time). On 12/3/04, the recalling firm issued a press release to alert the medical community of the problem and the need to return the device. The recalling firm issued a recall letter dated 12/7/04, to the Director of Purchasing, informing them of the adverse events associated with the use of the product and the need to return all of their inventory.

Root Cause

Other

Distribution Pattern

The products were shipped to through specialty dealers/distributors to medical facilities nationwide. The product is also sold to government accounts and sales representatives. The product is also sold in Taiwan, Brazil, Hong Kong, Africia, and Canada.