DRLock Volar Plate Screw System, Model #DVR-900. The product is a kit contained in an aluminum tray case that contains the following: IMPLANTS: 6 stainless steel plates (2-lefts, 2-rights, and 1-right long and 1-right left), 48 distal locking-screws (ranging from 10-24mm), 32 distal locking pegs, 36 proximal locking screws, 24 proximal non-locking screws. INSTRUMENTS: 2 Bending Pliers*, 1 Bone Reduction Forcep*, 2 Hohmann Retractors *, 1 Ratchet Handle*, 1 Periosteal Elevator*, 1 Dingman Elevator *, 1 K & K Elevator *, four Driver tips *, five drill guides*, 2-1.6 Drill Bits *, 2-2.3 Drill Bits *, 1 Plate Holder Verbruge *, 2 Depth Gages *, 6 K wires.

FDA Recall Z-0349-2007 · Posted 2007-01-19

Recall StatusTerminated
Product CodeHRS
Date Initiated2006-08-03
Date Terminated2012-09-10
Product Quantity180

Reason for Recall

The 48 distal locking-screws (ranging from 10-24mm) contained in the kit are not within established specifications, specifically the minor diameter of the screws are undersized.

Firm Action

The recalling firm contacted all consignees by phone on or about 8/3/06.

Root Cause

Other

Distribution Pattern

The product was distributed to 5 consignees located in the U.S.: Archway Medical, Columbia, SC; Gulf Coast Surgical, Stafford, TX; Vaugham Medical, Wheat Ridge, CO; Belemy Medical, Cleveland, OH; and Crystal Clinic, Akron, OH.