A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis
FDA Recall Z-0349-2023 · Posted 2022-12-01
Recalling FirmPrimus Corporation dba Trinity Biotech
Recall StatusTerminated
Product CodeJPK
Date Initiated2021-10-05
Date Terminated2023-08-09
Product Quantity340 kits
Reason for Recall
The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.
Firm Action
The firm sent an Urgent Filed Safety Notice their consignees on 10/05/2021 by mail. The notice explained the issue, risk and requested the destruction of of the affected devices.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Turkey, India, Romania, Thailand, Vietnam, S. Korea, Peru, Pakistan Austria, UAE, Chile, Italy.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source