Fellows "Mighty 8" Power strip (Part Number 04175-0008) and TrippLite Power strip (PS-415-HGULTRA - Part Number 13-00106-00) are two components sold as part of the Merge Hemo System.
FDA Recall Z-0350-2018 · Posted None
Advertisement
Recalling FirmMerge Healthcare, Inc.
Recall StatusTerminated
Product CodeDQK
Date Initiated2017-04-06
Date Terminated2019-07-10
Product Quantity1,150 power strips
Reason for Recall
Merge is conducting a hardware recall (replacement) of Fellows "Mighty 8" Power Strip and TrippLite Power Strip due to overheat of input resistor causing the power strip to smoke.
Firm Action
Merge began notify customers via an Urgent Medical Device Recall letter on 4/6/2017 via mail. The letter identified affected product, stated the issue, and informed customers that Merge would replace affected product. A response form is requested to be returned to Merge. Questions can be sent via email to [email protected]
Root Cause
Component design/selection
Distribution Pattern
Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.