Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Product List Number 16026

FDA Recall Z-0353-2017 · Posted None

Recalling FirmHospira Inc.
Recall StatusTerminated
Product CodeFRN
Date Initiated2013-02-05
Date Terminated2016-12-15
Product Quantity20,311

Reason for Recall

Hospira has received reports of Malfunction S205 Backup Battery Failure alarms in Symbiq v3.13 pumps. This issue to an incorrect installation of components on some of the Power Supply Controller (PSC) Boards distributed prior to September 2012. On pumps where incorrect installation of components has occurred, the backup battery power discharges and subsequently, the battery cannot be charged. Because of the battery depletion, or lack of charge, a visual and audible warning alarm, S205, is displayed. An S205 alarm is intended to notify the customer that the coin call battery, powering the backup buzzer, may need to be charged or replaced.

Firm Action

Hospira sent an Urgent Field Device Correction letter dated Feburay 5, 2013, to all affected customers. Hospira has implemented a correction to the manufacturing process to address this issue. Symbiq v3.13 pumps that have been shipped from Hospira since September 2012 and pumps that were upgraded to v3.13 in the Field after September 2012 are not impacted by this issue. If you are unsure of when you were shipped Symbiq v3.13 pumps or when your devices were upgraded, contact Hospira. Please complete the attached reply form and return it via fax number or e-mail address on the form, even if you do not currently have the impacted infuser. lease contact Stericycle at 1-866-201-9072 (M-F, Bam - 5pm EST) to obtain additional copies of the reply form. If you have further distributed these devices, please notify your accounts who may have received these infusers from you and ask them to contact Stericycle at 1-866-201-9072 (M-F, 8am- 5pm EST) to receive a reply form.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution and Canada