Siemens brand PRIMUS Linear Accelerators equipped with the optional Retractable Beam Shield and MEVATRON Linear Accelerator with optinal Retractable Beam Shield.

FDA Recall Z-0355-06 · Posted 2006-01-06

Recall StatusTerminated
Product CodeIYE
Date Initiated2003-01-08
Date Terminated2006-05-15
Product Quantity125 units

Reason for Recall

Incomplete installation of the optional Beam Shield may allow unexpected doses of radiation outside the radiation protected areas.

Firm Action

On 1/8/03, the firm initiated the recall and its notification was via letters informing its customers of the corrective action.

Root Cause

Other

Distribution Pattern

28 units shipped within the US, 97 shipped internationally, to countries including Japan, India, Taiwan, Hungary, Italy, France, Colombia, Germany, Austria, CAnada, Croatia, Poland, Croatia, Brazil, Turkey, Greece, Egypt, China, South Korea, Malaysia, Herzogovina.