Terumo Advanced Perfusion System 1 (6 inch diameter Roller Pump; Model 801041). Extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures.

FDA Recall Z-0355-2009 · Posted 2008-12-11

Recall StatusTerminated
Product CodeDTQ
Date Initiated2008-07-18
Date Terminated2009-07-07
Product Quantity2,599

Reason for Recall

Overspeed and underspeed errors, erratic pump behavior, jerky operation at low RPMs, pump instability, pump slowdowns and pump stops due to overspeed max events may occur due to a grease leak onto the motor speed encoder disk.

Firm Action

Consignees were notified of the problem via recall letter dated 7/18/08. The firm began conducting field upgrades on the roller pumps in 07/08. For questions or concerns, contact Terumo CVS Customer Service at 1-800-521-2818.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution including USA and countries of Australia, Belgium. Canada, Columbia, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.