ProTrack Microcatheter. Model/Catalog Numbers CIC38-145, CIC35-145. Sterile, single use device intended to be used as a standard infusion wire or as an exchange guide for exchanging guidewires. Coaxial Injectable catheter guide wire. Cardiovascular.

FDA Recall Z-0356-2016 · Posted 2015-12-01

Recalling FirmBaylis Medical Corp *
Recall StatusTerminated
Product CodeDQX
Date Initiated2015-10-28
Date Terminated2016-11-17
Product Quantity433 units

Reason for Recall

Microcatheter may have circumferential defects (cracks) along its shaft.

Firm Action

Recall notification letter was sent to affected consignees on 10/28/2015 via courier. The letter contained a fax-back form to be completed by the consignee and sent to the quality department of the recalling firm.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution.