Fibrin Analysis Catheter Testing System includes the component BD Vacutainer Luer-Lok Access Device; Catalogue #364902 as a finished sterile device, Non Pyrogenic in unopened undamaged package, Disposable/ Single use/ Do Not Reuse; for use with multi lumen; (4 Venous Catheter sizes) 20 cm. 12 fr. Central; 16 cm. 7-7.5-8 fr. Central; 20 cm 7.7.5-8 fr Central; 16cm 12 fr. Central. Manufactured in the USA for: CRS Medical Diagnostics, Inc., Pewaukee, WI.
FDA Recall Z-0357-2007 · Posted 2007-01-18
Advertisement
Recalling FirmContinental Medical Labs,Inc
Recall StatusTerminated
Product CodeGCJ
Date Initiated2006-10-18
Date Terminated2007-01-18
Product Quantity840 kits
Reason for Recall
Fibrin Analysis Catheter Testying System contains the ''Luer Lok Access Device'' that fails to lock securely to certain catheter devices because the male luer taper surface is not within the specification of the taper.
Firm Action
Consignee was notified by certified mail on 10/19/06.
Root Cause
Process control
Distribution Pattern
Product was distributed to one consignee located in Wisconsin
Official Source