IOL Master 5.5: Software versions 7.5 and 7.7; Ophthalmic: used to obtain ocular measurements and perform calculations to allow physicians to determine appropriate IOL power and type for implantation.
FDA Recall Z-0357-2016 · Posted 2015-12-01
Advertisement
Recalling FirmCarl Zeiss Meditec AG
Recall StatusTerminated
Product CodeHJO
Date Initiated2015-10-27
Date Terminated2016-06-06
Product Quantity49
Reason for Recall
IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
Firm Action
Urgent Medical Device Correction letters will be mailed on Friday, November 13, 2015 as the IOL SW update material is now available. Letter informs customers of the issue and provide instructions for identifying the software version and for installing the software update with illustrations.
All customers are requested to complete and return the confirmation form to Stericycle as soon as the upgrade is complete or it has been determined that your unit is not running the affected version of the software.
Root Cause
Software design
Distribution Pattern
Nationwide Distribution.