2.4mm x 2.5 Fluted Pin Pouch; Catalog Number: 351-90-22; Orthopedic surgical instrument used in the implantation of the Vantage Total Ankle System - Product Usage: To assist the surgeon in the implantation of Vantage Total Ankle System components according to a conventional technique for total ankle replacement. These instruments are single-use, provided sterile, and intended for transient use.
FDA Recall Z-0361-2020 · Posted None
Advertisement
Recalling FirmExactech, Inc.
Recall StatusTerminated
Product CodeNDL
Date Initiated2019-10-02
Date Terminated2023-05-19
Product Quantity9 units
Reason for Recall
Potential for the 2.4mm x 2.5 Fluted Pin Pouch to contain a single pin rather than four (4) as intended.
Firm Action
Exactech notified customers on about 10/02/2019, via "URGENT MEDICAL DEVICE RECALL" letter. Customers were informed that the 2.4mm x 2.5" Fluted Pin Pouch was found to contain a single pin rather than four (4) as labeled.
Instructions included to immediately cease distribution or use of the affected product, provide the information to customers if further distributed, identify and quarantine any of the affected devices in inventory, and complete and return the provided Recall Acknowledgement Form and Recall Inventory Response Form to Exactech via email to [email protected].
Root Cause
Error in labeling
Distribution Pattern
US Nationwide distribution in the states of GA, NY, OH, VA.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.