PICIS, CareSuite Family- OR Manager (4235), Preop Manager (4250), Anesthesia Manager (4220), PACU Manager (4245) & Critical Care Manager (1110), patient information systems.

FDA Recall Z-0365-2010 · Posted 2009-11-18

Recalling FirmPicis Inc.
Recall StatusTerminated
Product CodeNSX
Date Initiated2009-04-08
Date Terminated2014-12-22
Product Quantity325

Reason for Recall

Demographic data, most notably allergy and precaution data, can be overwritten with incomplete data or blanks by the interface between non-Picis clinical systems (your "HIS" or "CIS") and certain Picis applications, if that interface is not configured properly.

Firm Action

On April 8th 2009, Picis initiated a customer notification letter to the installed base of potentially affected users, explaining the potential source of patient data conflict and to provide support in choosing the appropriate configuration option and implementation. The notification letters included with this report package contain additional details regarding configuration options.

Root Cause

Software design

Distribution Pattern

Worldwide distribution: USA, Australia, Austria, Belgium, Denmark, Finland, France, Germany, Italy, Malaysia, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, and United Kingdom