Patient Support Material Number: 9890 010 87431, used with:EasyDiagnost ; Model: 706050

FDA Recall Z-0365-2016 · Posted None

Recall StatusTerminated
Product CodeKPR
Date Initiated2014-08-19
Date Terminated2016-06-14
Product Quantity260 units, not all units have the patient support accessory

Reason for Recall

The hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.

Firm Action

The Field Safety Notice FCO70600086 will be mailed on August 19, 2014. Philips Medical Systems will send out certified letters and will have the receipts returned for tracking purposes as well as have a Field Service Engineer visit each site to provide the fix when the Field Change Order is released.

Root Cause

Device Design

Distribution Pattern

Distributed in the states of MN, KY, MA, VA, TX, NC, HI, FL, AZ, CT, and WY and the countries of Argentina, Australia, Austria, Canada, Denmark, Germany, Netherlands, Russia, Saudi Arabia, Singapore, Spain, and Switzerland.