Terminated

Zimmer ZMR Femoral Spout & Cone Bodies and Trabecular Metal Primary Hip Prosthesis Femoral Stem SIZE 16 171 MM STEM LENGTH 12/14 NECK TAPER - STANDARD BODY - STANDARD NECK OFFSET TIVANIUM TI-6AL-4V ALLOY/TANTALUM STERILE. Item number: 00-7864-016-00

FDA Recall Z-0366-2012 · Posted None

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2011-11-16
Date Terminated2013-08-28
Product Quantity2 (1 device is domestic)

Reason for Recall

Lack of assurance of proper metal fatigue strength due to metal grain structure anomaly. In 2006 a recall was initiated as a result of the grain anomaly; however, these 4 lots had been split off during normal production and had not been included in the 2006 recall.

Firm Action

On 11/16/11, Zimmer sent letters titled "URGENT MEDICAL DEVICE RECALL" to distributors via electronic mail and to risk managers and surgeons via FedEx indicating that the product must be immediately removed from use and quarantined. " Stop using the affected lots and place affected product in quarantine. " Review available product on the shelf " Identify those part/lot combinations listed in the recall communication " Immediately remove any implants found that match part/lot combinations listed in the recall communication and quarantine " Return affected lots and signature of acknowledgement to Zimmer

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution: USA (NY), AUSTRALIA, SWITZERLAND, and DENMARK