RANDOX NEFA (Non-Esterified Fatty Acids) FA115 Lot 485343, GTIN: 5055273203066.

FDA Recall Z-0367-2020 · Posted None

Recall StatusTerminated
Product CodeJLG
Date Initiated2019-09-23
Date Terminated2020-05-07
Product Quantity212 (19 US; 193 OUS)

Reason for Recall

Non-Esterified Fatty Acids (NEFA) FA115 batch 485343 is not meeting the performance claims and should not be used in any further testing.

Firm Action

On September 23, 2019 the firm distributed Urgent Medical Device Recall letters by e-mail instructing them to: - Discontinue use of and discard any of the product detailed above immediately or return remaining product to Randox Laboratories. - Review results generated with the affected batches in line with the clinical profile of the patient. - Discuss the contents of this notice with your Medical Director. - Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. - Complete and return the response form 12187-QA to [email protected] within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization. On October 8, 2019 the firm distributed an amended Urgent Medical Device Recall. However, these letters were not identified as an Amended or Corrected notice. The notice is also still dated 09/23/2019. This updated notice identified the expiration date for the affected product as 28Mar2021. On October 16, 2019, the firm sent a second amended notice to customers via email.

Root Cause

Nonconforming Material/Component

Distribution Pattern

WV (USA), International: GB, BE, CZ, FR, GR, IE, IT, NL, PL, SK, ES, CH, TR, AU, BR, MX, RU, SG, TW