GE Healthcare OEC 9900 mobile fluoroscopic x-ray system

FDA Recall Z-0368-2008 · Posted 2008-09-29

Recall StatusTerminated
Product CodeJAA
Date Initiated2007-12-14
Date Terminated2012-04-13
Product Quantity710

Reason for Recall

Possible overexposure: Failure to apply the entrance exposure rate (EER) tube current limit calibration to the automatic exposure rate (AER) control system when the anatomical profile mode is changed from the default selection to another selection. Manual technique modes are not affected.

Firm Action

GE OEC sent a letter to customers December 15, 2007, and scheduled field representative visits for corrective software solution installation.

Root Cause

Software design

Distribution Pattern

Worldwide distribution: USA, Finland, Netherlands, Germany, Switzerland, France, United Kingdom, Australia, Italy, Ireland, Spain, Sweden, Belgium, New Zealand, China, Singapore, Poland, Norway, and India.