Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

PROLARYN PLUS Injectable Implant 1.0 cc Plus Calcium Hydroxlyapatite (CaHA) Injectable Implant, Model No. 8044M0K5. An injectable implant used to treat vocal fold insufficiency.

FDA Recall Z-0368-2016 · Posted 2015-12-03

Recall StatusTerminated
Product CodeKHJ
Date Initiated2015-10-26
Date Terminated2017-05-08
Product Quantity9,439 units

Reason for Recall

The expiration date encoded in the UDI barcode on the label on the product carton and the peelable patient identification label is incorrect.

Firm Action

Letters were issued 11/4/2015 to consignees. The letter informs the consignees that they can continue to safely use the affected products by referring to the USE BEFORE date. They letter does not require for any product to be returned. Also, there is no sub-recall required as the product was not distributed to distributors or wholesalers.

Root Cause

Other

Distribution Pattern

Distributed US (nationwide) including the Virgin Islands and South Africa.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑