Terminated

IMx Folate Medium Control, list 2220-11; the pack contains 8 - 1 mL bottles of IMx Folate Medium Control M 7.0 ng/mL; Abbott Laboratories, Abbott Park, IL 60064 USA

FDA Recall Z-0369-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeCGN
Date Initiated2003-12-16
Date Terminated2004-07-20
Product Quantity54 packs

Reason for Recall

The is an increase in the number of control values out of range low, and diluted patient specimens may yield lower than expected results.

Firm Action

Abbott notified their international affiliates via e-mail on 12/16/03 and 1/5/04. Device recall letters dated 12/16/03 and 12/31/03 will be sent to the end users by the affiliates, informing them of the unapproved serum albumin used to manufacture the folate products, and listing the impact to performance for each affected components. There are separate customer recall letters for each assay platform -- ARCHITECT, AxSYM and IMx.

Root Cause

Other

Distribution Pattern

There was no U.S. distribution. The products were shipped internationally through Abbott affiliates in Canada, Germany, Singapore, Hong Kong, Australia, Brazil, New Zealand, Columbia, Chile, Uruguay, United Kingdom, Venezuela, Dominican Republic, Argentina and Curacao.