UniCel Dxl 800 Access Immunoassay Systems (Reaction Vessels).
FDA Recall Z-0371-05 · Posted 2004-12-28
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeJJE
Date Initiated2004-10-29
Date Terminated2005-03-16
Product Quantity56 boxes
Reason for Recall
Beckman Coulter has confirmed reports of rare occurrences of UniCel Dxl 800 Access Immunoassay Systems Reaction Vessels that are deformed or flattened in their packaging bags for lot 04250160. Deformed or flattened RVs could present a rish of causing jams on the UniCel Exl 800 Access Immunoassay Systems that may require service intervention and could compromise sample analysis.
Firm Action
All customers that were shipped lot 042500160 of the reaction vessels were contacted via telephone (or via on-site visit - Canada only) October 29 - November 1, 2004.
A Product Correction Action Letter dated November 2,2 004, sent via U.S. mail to all customers, once contacted by phone with fax back forms.
Root Cause
Other
Distribution Pattern
FL, HI, IL, MS, MI, NC, NY, OH, OK, TX, UT, VA, WI, and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.