Valeant Pharmaceuticals North America Model No. Spag-2 Small Particle Size Aerosol Generator SPAG-2 6000 is a pneumatic flow system indicated for the administration of Viazole uses a nebulizer to generate a fine aerosol of hydrated Virazole
FDA Recall Z-0371-2016 · Posted 2015-12-04
Advertisement
Recalling FirmValeant Pharmaceuticals North America LLC
Recall StatusTerminated
Product CodeCAF
Date Initiated2015-09-14
Date Terminated2017-06-27
Product Quantity61 units
Reason for Recall
SPAG-2 6000 Lot J3522070 was released to commerce without being tested for particle size per quality control procedure SPAG-2515.
Firm Action
Valeant Pharmaceuticals International issued recall letter/return response form dated 9/11/2015 via 2nd day air service Customers were requested to check their inventory for the SPAG-2 device and sgregate it for return shipment.
Root Cause
Under Investigation by firm
Distribution Pattern
AL, AZ, CA, CO, IL, OH, MO, NC, NY, PA., SC, UT and TX