EVO Excel, Emission Computed Tomographic System, Material Number 10910501
FDA Recall Z-0372-2019 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc.
Recall StatusTerminated
Product CodeKPS
Date Initiated2018-10-05
Date Terminated2021-02-11
Product Quantity264 units
Reason for Recall
The Evo Excel External-ECG Power-Connection current limit is mis-labeled as 1.0A. It should be labeled 0.55A.,
Firm Action
Siemens disseminated the notices by letter on 10/05/2018. Firm will deliver a Symbia Evo Excel Addendum and DVD addendum with this updated information at a later date.
Root Cause
Other
Distribution Pattern
Worldwide