Warm Air Model 135 Warming Units.

FDA Recall Z-0374-05 · Posted 2004-12-30

Recall StatusTerminated
Product CodeBTE
Date Initiated2004-11-05
Date Terminated2008-10-29
Product Quantity649 units.

Reason for Recall

Burning smell, burned microprocessor board. The firm is modifying the connectors and wiring gauge configuration of the device.

Firm Action

The firm notified consignees by certified letter on November 5, 2004.

Root Cause

Other

Distribution Pattern

The product was distributed throughout the United States and to foreign countries including Saudi Arabia, Taiwan, Japan, and Canada.