GEM PREMIER 3000

FDA Recall Z-0378-06 · Posted 2006-01-11

Recall StatusTerminated
Product CodeCGA
Date Initiated2005-12-16
Date Terminated2007-05-21
Product Quantity2021 units

Reason for Recall

Software error may report higher results for glucose and lactate

Firm Action

Instrumentation Laboratory notified customers by letter dated December 2005 on 12/16/2005. Customers were notifed if the error message occurs the lab should diable the glucose and lactate channels or remove cartridge. A software upgrade will be released in early 2006.

Root Cause

Other

Distribution Pattern

Nationwide Foreign: Canada