GEM PREMIER 3000
FDA Recall Z-0378-06 · Posted 2006-01-11
Advertisement
Recalling FirmInstrumentation Laboratory Co.
Recall StatusTerminated
Product CodeCGA
Date Initiated2005-12-16
Date Terminated2007-05-21
Product Quantity2021 units
Reason for Recall
Software error may report higher results for glucose and lactate
Firm Action
Instrumentation Laboratory notified customers by letter dated December 2005 on 12/16/2005. Customers were notifed if the error message occurs the lab should diable the glucose and lactate channels or remove cartridge. A software upgrade will be released in early 2006.
Root Cause
Other
Distribution Pattern
Nationwide Foreign: Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.