Roche Ammonia/Ethanol/CO2 Control A (Abnormal) for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas c111 and cobas 46250 c 501 clinical chemistry analyzers; Cat. Number. 20753009190. Roche Diagnostics Corporation, Indianapolis, IN

FDA Recall Z-0379-2008 · Posted 2007-12-29

Recall StatusTerminated
Product CodeJJY
Date Initiated2007-11-14
Date Terminated2009-09-22
Product Quantity1,296

Reason for Recall

High Quality Control Recovery: Undetected high quality control recovery in bicarbonate assay.

Firm Action

Consignees were notified of this recall by an Urgent Medical Device Recall letter sent on 11/14/07. The letter instructed the users to immediately discontinue use of Ammonia/ Ethanol/CO2 Control N and Ammonia/Ethanol/CO2 Control A (with listed lots) for quality control assessment of the bicarbonate assay. They were informed that controls may still be used to assess ammonia and ethanol, but alternative controls must be used for assessment of the bicarbonate assay. Consignees are instructed to destroy products at their facility. Consignees were notified of the availability of replacement product via letter dated .

Root Cause

Process design

Distribution Pattern

Nationwide Distibution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.