Product Brand Name(s): Alere Cholestech LDX Lipid Profile"GLU Test Cassette, Model 10-991. Product Generic Name: Alere Cholestech LDX Lipid Profile"GLU Test Cassette. Description of the product: Lipid. Glucose panel Test cassettes. Product packaging: 10 individually pouched cassettes in a labeled kit box. Intended for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides, and/or glucose in whole blood, serum, and plasma.

FDA Recall Z-0379-2014 · Posted 2013-11-22

Recalling FirmAlere San Diego, Inc.
Recall StatusTerminated
Product CodeCGA
Date Initiated2013-10-23
Date Terminated2014-09-25
Product Quantity3,621 kits

Reason for Recall

Alere initiated a recall of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 320167, becuase an incorrect lot number was printed on the outer 10 pack kit box label. The incorrect Lot # number printed was 321067.

Firm Action

Alere initiated this recall by sending out notification letters to customers via fax, email, or direct mail. The letter titled "URGENT MEDICAL DEVICE NOTICE", dated October 23, 2013 is accompanied by a response form/Verification form. The notification informed customers of the reason for recall, product description with codes, required actions by the customers, and contact information.

Root Cause

Labeling Change Control

Distribution Pattern

Worldwide Distribution: USA (nationwide) and Internationally to Canada.