Terminated

ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.

FDA Recall Z-0379-2017 · Posted 2016-11-09

Recalling FirmAbbott Laboratories
Recall StatusTerminated
Product CodeKHP
Date Initiated2016-10-11
Date Terminated2017-05-17
Product Quantity26, 837 kits

Reason for Recall

Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).

Firm Action

Abbott Laboratories sent an Product Correction Immediate Action Required letter dated October 10, 2016, to all affected customers. Please review this letter with your Medical Director Taking the above information into consideration, you may continue to use ARCHITECT Lactic Acid reagent (LN 9D89-21) Please complete and return the Customer Reply Form If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter Please retain this letter for your laboratory records. If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service. For further questions please call (847) 937-2550.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide and Worldwide Distribution.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑