Terminated

Endoform Dermal Template 4x5, SKU 529313 Product Usage: Endoform Dermal Template is a sterile, single use ovine forestomach derived extracellular matrix intended to cover, protect and provide a moist wound environment.

FDA Recall Z-0379-2019 · Posted 2018-11-02

Recalling FirmAROA BIOSURGERY
Recall StatusTerminated
Product CodeKGN
Date Initiated2018-11-27
Date Terminated2019-04-09
Product Quantity8,853 total boxes

Reason for Recall

Potential for pouch seal failure

Firm Action

The firm sent a Customer Notification letter dated November 27, 2018. The letter identified the affected product, problem and actions to be taken. Customers are encouraged to visually inspect the seals on devices from these lots prior to use with particlar care. Should you observe a comprimised seal, please discard the device and select a new, fully sealed device for use.

Root Cause

Employee error

Distribution Pattern

US Nationwide Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑