Terminated

RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

FDA Recall Z-0380-2018 · Posted None

Recall StatusTerminated
Product CodeMUJ
Date Initiated2017-07-17
Date Terminated2018-08-29
Product Quantity117 units

Reason for Recall

For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.

Firm Action

Ray Search sent a Field Safety Notice letter dated July 14, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter included a response form with instructions for workarounds until the software is corrected.

Root Cause

Device Design

Distribution Pattern

US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.