Medication Bottle (U22-3), a component of the MicroAir Vibrating Mesh Nebulizer, Model NE-U22V/NE-22VAC; Distributed by: Omron Healthcare, Inc., 300 Lakeview Parkway, Vernon Hills, IL 60061
FDA Recall Z-0382-05 · Posted 2005-01-05
Advertisement
Recalling FirmOmron Healthcare, Inc.
Recall StatusTerminated
Product CodeCAF
Date Initiated2004-11-02
Date Terminated2007-06-26
Product Quantity63,560 units
Reason for Recall
A potential failure of the vibrating mechanism of the medication bottle which causes a failure to transmit vibration and results in no nebulization.
Firm Action
Omron sent 'Removal Notices' dated 11/2/04 to all of their direct accounts receiving the nebulizers during the affected time frame. The accounts were informed of the potential failure of the vibrating mechanism, and listed the serial numbers of the nebulizers that may contain the affected medication bottles. Medication bottles with either a blue dot near the bottle lever or a blue O-ring around the body of the bottle were manufactured after corrections were made and are not subject to the recall. Affected product should be returned to Omron for replacement. Any questions were directed to Customer Service at 800-231-4030.
Root Cause
Other
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.