Terminated

Tri-Nex Healing Abutment, 0.50 mm x 5 mm height, Catalog No. HA-L-50-5. The NSI Implant System is intended to be implanted in the upper and lower jaw arches to provide support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses, or full arch prostheses. It further adds the option for immediate loading on single and splinted multiple unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function.

FDA Recall Z-0382-2014 · Posted 2013-11-22

Recall StatusTerminated
Product CodeNHA
Date Initiated2013-09-13
Date Terminated2014-01-21
Product Quantity197 units

Reason for Recall

Southern Implants is recalling the Tri-Nex healing abutment because the package was incorrectly labeled and packaged as the Tri-Nex healing abutment, 05.0 mm x 5 mm height (HA-L-50-5). The package contains the Tri-Nex healing abutment 05.0 mm x 6.0 mm Flare x 5 mm Height (HA-L-50W-5).

Firm Action

Southern Implants, Inc. sent an Urgent Product Action letter dated September 11, 2013, to all affected customers. The letter identiied the product, the problem, and the action to be taken by the customer. Customers were instructed to fill out the customer response form and fax/email the completed form to Customer Support at (866) 903-9272 or [email protected]. For questions regarding this recall call 866-903-9272.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally to Canada, Guatemala, Taiwan.