"***SENTINEL***Cannulated Drill Bit***S8580***8 X 229mm (9")***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 33773-4908 USA***LOT***2012-03***". This drill bit is an eccentric mono-fluted cutting edge for drilling bone tunnels in ACL and PCL reconstruction.
FDA Recall Z-0383-2013 · Posted 2012-11-19
Advertisement
Recalling FirmLinvatec Corp. dba ConMed Linvatec
Recall StatusTerminated
Product CodeHTW
Date Initiated2012-04-10
Date Terminated2013-11-21
Product Quantity70 total
Reason for Recall
Linvatec Corporation dba ConMed Linvatec in Largo, FL is recalling the S8580, 8.0MM SENTINEL Cannulated Drill Bit, lot numbers 322225 and 322226; and the S8585, 8.5MM SENTINEL Cannulated Drill Bit, lot number 322287. Recall was initiated due to the incorrect labeling on package.
Firm Action
ConMed/Linvatec sent an Urgent Medical Device Recall Notificaiton letter via phone and FedEx on April 9, 2012 and again on May 11, 2012, to all affected customers. They also sent the same Notification on May 14, 2012, to international accounts. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete a reply formand fax it to 727-319-5701. Customers were also instructed to return all affected product to the firm. All returned product under this removal action will be identified, segregated and quarantined per firm procedures, etc.
Customers with questions should contact Customers Service at [email protected].
For questions regarding this recall call 727-392-6464.
Root Cause
Labeling Change Control
Distribution Pattern
Worldwide Distribution - USA including NJ, UT, GA, NY, PA, IL, LA, OH, CA, MS and Internationally to Canada, Japan, and Romania.