PriMatrix is an acellular dermal tissue matrix derived from fetal bovine dermis. The device is supplied sterile in a variety of sizes to be trimmed by the surgeon to meet the individual patient's needs.
FDA Recall Z-0383-2018 · Posted None
Advertisement
Recalling FirmTEI Biosciences
Recall StatusTerminated
Product CodeKGN
Date Initiated2017-06-15
Date Terminated2018-03-15
Product Quantity4081 units
Reason for Recall
During Final Assembly Operation, it was noticed that some of the PriMatrix Instructions for Use (IFUs) from Lot RM1608011 Part 520-002v10 contained extra pages and/or missing pages:
" 22 IFUs contained duplicate pages 15 - 26 & 51 - 62 and were also missing pages 27 - 50
" 3 IFUs were missing pages 27 - 50
" 1 IFU was missing pages 1 - 2 & 75 - 76
" 1 IFU was missing pages 15 - 62
While pages 27 50, 75 76, and 15 62 contain instructions written in a language other than English, pages 1 2 contain information regarding the indications for use, contraindications, device description, warning and precautions, and potential complications of the device.
Firm Action
Integra LifeSciences Corporation notified (all U.S.) customers via Letter/Medical Device Correction Field Safety Notice dated June 15, 2017. Letter delivered by courier service, facsimile or email July 15, 2017.
Root Cause
Error in labeling
Distribution Pattern
U.S.