Single Analyte Urine DAU Control (5 m L Norbuprenorphine, Level II control For In Vitro Diagnostic Use Only. store at 2 - 8 degrees C. Contents: Contains 13 mg/Ml of Norbuprenorphine in human urine with sodium azide as a preservative. Manufacturer: Lin-Zhi International, Inc. 670 Almanor Ave. Sunnyvale, CA 94085 The Norbuprenorphine Drug of Abuse (DUB) Positive control is used as assayed quality control material to monitor the precision of the Lin-Zhi International, Inc. Buprenorphine Enzyme immunoassay on the Synchron clinical analyzer. It in intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine.

FDA Recall Z-0386-2014 · Posted 2013-11-25

Recall StatusTerminated
Product CodeLAS
Date Initiated2013-10-03
Date Terminated2013-12-26
Product Quantity60 items

Reason for Recall

Product catalog # A68825 Norbuprenorphine 13 ng/mL Level 2 Controls gives a decreased concentration reading due to the degradation of the NBUP analyte in solution.

Firm Action

Beckman Coulter sent a Manufacturer Field Action Letter dated November 2013, to the one distributor. The letter identified the product the problem and the action needed to be taken by the customer. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. If you have any questions regarding this notice, please contact our Customer Support Center Via our website, http://www.beckmancoulter.com/customersupport/support Via phone, call 1-800-854-3633 in the United States Outside the United States, contact your local Beckman Coulter Representative. We apologize for any inconvenience.

Root Cause

Under Investigation by firm

Distribution Pattern

US Distribution in CA.