LOCIQ Book System models: 2399921, 2399923, 2349933, 2349934 and 2349935
FDA Recall Z-0387-06 · Posted 2006-01-11
Advertisement
Recalling FirmGeneral Electric Med Systems LLC
Recall StatusTerminated
Product CodeIYN
Date Initiated2004-09-02
Date Terminated2007-01-05
Product Quantity2698
Reason for Recall
The Estimated Fetal Weight (EFW) result will be incorrect if the operator fails to perform the 'end exam' or new patient registration step before starting an EFW measurement procedure on the next patient.
Firm Action
Devices will have software upgrades done by GE Field Service Technicians. The upgrade will include a dialog box which will display on the screen when prior data exists and a new patient is being registered.
Root Cause
Other
Distribution Pattern
nationwide and worldwide