Phadia 1000 Instrument, Article Number 12-3800-01.
FDA Recall Z-0387-2018 · Posted None
Advertisement
Recalling FirmPhadia US Inc
Recall StatusTerminated
Product CodeDHB
Date Initiated2017-07-05
Date Terminated2018-10-04
Product Quantity70 units
Reason for Recall
The "Retry" command does not function properly which could cause a shortage of Wash and Rinse solution and affect assay performance and test results.
Firm Action
The firm sent letters with a return response form to customers. The firm instructed customers to immediately choose the "stop" command if presented with the 7-101 Liquid Sensor Error. Results from the assay should not be approved. A revision of the current software will have a mandatory installation soon.
Root Cause
Device Design
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.