Elation, Class I Medical Device Prosthetic product, Component, External, Limb, Ankle/Foot Parts Numbers: ELP0023L/R-ELP0026L/R, ELP0023RBR, ELP0024RBR &, ELP0028RBR.
FDA Recall Z-0388-06 · Posted 2006-01-12
Advertisement
Recalling FirmOssur
Recall StatusTerminated
Product CodeISH
Date Initiated2005-10-21
Date Terminated2007-03-20
Product Quantity48 (23 in US)
Reason for Recall
After one year devices assembled with a new type of O-rings could exhibit enough wear to allow a gradual oil leak which would increase play in the ankle sub-assembly of the foot.
Firm Action
Telephone communication followed by notification letter mailed to end users (patients) and prosthetics workshops. A pre-paid package to return faulty components to Ossur included with mailed letters. New replacement feet will be shipped to customers.
Root Cause
Other
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.