Brilliance 64, (Code 728231) Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.

FDA Recall Z-0389-2018 · Posted None

Recall StatusTerminated
Product CodeJAK
Date Initiated2017-11-10
Date Terminated2020-11-05
Product Quantity1

Reason for Recall

Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

Firm Action

An Urgent - Field Safety Notice Medical Device Correction letter, dated November 1, 2017, was sent to customers. The letter identified the affected device, described the issue, and discussed the hazard involved. Users are to cease use immediately and contact their Philips Field Service Engineer, if any unusual noises are heard within the gantry, or if the patient support tabletop is in free float. Philips Field Service Engineers will correct the system by replacing the affected fasteners. If further info or support concerning this issue is needed, customers can contact their local Philips rep or the Customer Care Solutions Center at 1-800-722-9377.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution -- US, Argentina, China, Republic of Korea, Russian Federation, Slovakia, Turkey, and United Kingdom. 4 in US, (1 military), 14 foreign accounts