Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Synthes Matrix 5.5.mm Left and Right Coronal Bender Product Usage: The Matrix 5.5.mm Left and Right Coronal Bender are used together to bend the 5.5 Matrix Rods in the coronal plane, and is intended for stabilization of the thoracic, lumbar, sacral, and/or iliac spine through screw and/or hook and rod fixation.

FDA Recall Z-0391-2014 · Posted 2013-11-26

Recalling FirmSynthes USA HQ, Inc.
Recall StatusTerminated
Product CodeHXP
Date Initiated2013-06-22
Date Terminated2015-09-10
Product Quantity209

Reason for Recall

A complaint was received of the tips of the Matrix 5.5.mm Left and Right Coronal Bender breaking while bending a spinal rod component of the Matrix System.

Firm Action

Synthes sent an Urgent Notice: Medical Device Recall letter dated June 22, 2013 to users and sales reps. The letter identified the affected product, problem and actions to be taken. For questions call 610-719-5450.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution - USA (nationwide) including GA, FL, CA, UT, IA, AZ, OH and Internationally to Canada.

Official Source

View FDA Recall →