OR Table, Top U26 H V, Model # 1909793; OR Table Top U26 H V U, Model # 1909794; OR Table Top U24 H V, Model # 1909796; OR Table Top U24 H V U, Model # 1909797; OR Table Top U24 H V W, Model # 1909798; OR Table Top U14 H V, Model # 1909799 Product Usage: Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Patient transport on the operating table top from a patient transfer system to the operating theater or from the operating theater to a patient transfer system.

FDA Recall Z-0391-2017 · Posted 2016-11-10

Recall StatusTerminated
Product CodeJEA
Date Initiated2016-10-25
Date Terminated2017-02-06
Product Quantity4 (US) & 229 (Foreign)

Reason for Recall

Trumpf has identified six incidents of pinched cables.

Firm Action

Trumpf Medical sent an Urgent Medical Device Correction letter dated October 7, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken The letter indicated that a authorized service provider will inspect the table tops and make necessary corrections. A confirmation notice was included with the letter which is to be returned. For contact Trupmpf Medical Systems, Inc. at 888-474-9359.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of AZ and SC and the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Iceland, Italy, Norway, Qatar, Slovenia, Spain, Sweden, Switzerland