Synthes Spiked Washers 13.5/6.5 Spiked Washers 13.5/6.5 are intended to be used in ligament reattachment or fixation, specifically readaptation of torn or avulsed ligaments.
FDA Recall Z-0394-2014 · Posted 2013-11-26
Advertisement
Recalling FirmSynthes USA HQ, Inc.
Recall StatusTerminated
Product CodeHTN
Date Initiated2013-10-03
Date Terminated2015-12-18
Product Quantity3087
Reason for Recall
During a retrospective assessment, Spiked Washers 13.5/6.5 were discovered to be mislabeled as Spiked Washers 13.5/5.5.
Firm Action
SYNTHES sent an Urgent Notice: Medical Device Recall letter dated October 3, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed call Synthes at 1-800-479-6329 to obtain a Return Authorization Number, complete the Verification Section at the end of the letter by checking the appropriate box indicating affected product located. Return the Verification Form to:
Credit/Returns
Synthes
1101 Synthes Avenue,
Monument, CO 80132
Customers were also asked to return the Verification Form indicating they did not have the affected product also. All documents were to be returned to Synthes by:
Fax 1-877-270-5855
Scan/email: [email protected]
Customers with questions were instructed to call 610-719-5450 or contact their Synthes Sales Consultant
For questions regarding this recall call 610-719-5000..
Root Cause
Finished device change control
Distribution Pattern
Nationwide Distribution