Terminated

Boston Scientific brand iLab Ultrasound Imaging System, for intravascular use; Model Number(s): iLab120lNS, iLab2401NS, Distributed by and Manufactured by: Boston Scientific Corporation IVUS, Technology Center, 47201 lakeview Parkway, Fremont, CA 94538

FDA Recall Z-0395-2007 · Posted 2007-02-01

Recall StatusTerminated
Product CodeDQO
Date Initiated2006-12-01
Date Terminated2007-12-05
Product Quantity24 units

Reason for Recall

The product fails to comply with applicable electronic product performance standard, in that the electromagnetic energy is in excess of the labeled conformance standards for radiated emissions.

Firm Action

A customer notification was sent December 1, 2006 indicating the need for the correction.

Root Cause

Other

Distribution Pattern

Nationwide and Portugal.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.