Open, Classified

EXALT Model D Single-Use Duodenoscope: Outer box UPN# M00542420 & Outer box UPN# M0054242CE0.

FDA Recall Z-0396-2024 · Posted 2023-11-21

Recall StatusOpen, Classified
Product CodeFDT
Date Initiated2023-09-27
Product QuantityUS: 586 devices & OUS: 81 devices

Reason for Recall

Poor image quality due to fluid ingress in the lens.

Firm Action

Notification of Urgent Medical Device Product Removal was issued via FedEx on Oct 3, 2023. The notice advises customers to immediately discontinue use and segregate affected product, and to complete and return the Reply Verification and Tracking Form (RVTF) even if you do not have any affected product. Customers are to post the notice near any affected product and notify end users. If further distributed, notify any customers to which product has been further distributed to. Reach out to your local Boston Scientific representative with any questions. The notice is batch / lot specific. Boston Scientific has implemented corrective changes and will make replacement EXALT Model D scopes available as soon as possible.

Root Cause

Process control

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA VT, WA, WI & WV. The countries of Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Great Britain, Italy, New Zealand, Spain, Sweden, Switzerland.