Terminated

VKS Knee System; VKS Distal Femoral Resection Guide, Model Number 1214122, Revisions A and B. (Re-usable surgical instrument)

FDA Recall Z-0397-06 · Posted 2006-01-18

Recalling FirmPlus Orthopedics USA
Recall StatusTerminated
Product CodeJWH
Date Initiated2005-12-02
Date Terminated2007-01-29
Product Quantity14

Reason for Recall

The adjustment knob tower on one device was found to have broken off from the body of the instrument. This did not happend during a surgery and is believed to have been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.

Firm Action

The adjustment know tower on one device was found t have broken off from the body of the instrument. This did not happend during a surgery, and is belived tohave been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.

Root Cause

Other

Distribution Pattern

Arizona, California, Colorado, Idaho, New York, Oklahoma, Texas, Utah, Wisconsin