VirtuTRAX Instrument Navigator, Sensor with 10 17 GA VirtutTRAX

FDA Recall Z-0397-2022 · Posted None

Recall StatusTerminated
Product CodeIYO
Date Initiated2021-10-08
Date Terminated2024-09-30
Product Quantity18 boxes (5 units/box)

Reason for Recall

There is potential that the VirtuTRAX device will not tighten onto the instrument to be used.

Firm Action

The firm distributed an "URGENT: MEDICAL DEVICE RECALL" notice on 10/21/2021 by email. The notice explained the problem and requested acknowledgement of the notice. Entities further distributing the products were directed to notify those to whom the devices were distributed. The firm will be replacing the units.

Root Cause

Device Design

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Belgium, Canada, China (Mainland), China (Taiwan), Denmark, France, Germany, Italy, Japan, Netherlands, Singapore, South Korea, Spain, Thailand.