Terminated

Triathlon X3 CR Tibial Bearing Insert for knee prosthesis Size #2 - 11MM Ref # 5530-G-211 2011-09 Howmedica Osteonics Corp.

FDA Recall Z-0399-2007 · Posted 2007-02-08

Recall StatusTerminated
Product CodeJWH
Date Initiated2006-12-04
Date Terminated2007-02-21
Product Quantity23 units

Reason for Recall

A product mix-up in which the labeling indicates a Triathlon X# CR Tibial Bearing Insert Size #2 - 11MM, however, the device inside the package is actually a Triathlon X3 CR Tibial Bearing insert Size # 7 - 11MM and vice versa.

Firm Action

Notification letters and Product Accountability forms were sent on 12/8/2006 by Federal Express with return receipt.

Root Cause

Other

Distribution Pattern

Nationwide and Canada.