ARCHITECT STAT Myoglobin Calibrators; list 02K43-01; each kit contains 6 bottles of calibrators ranging from 0 ng/mL to 1200 ng/mL; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064
FDA Recall Z-0400-06 · Posted 2006-01-18
Recalling FirmAbbott Laboratories MPG
Recall StatusTerminated
Product CodeDDR
Date Initiated2005-10-17
Date Terminated2006-05-25
Product Quantity11 kits
Reason for Recall
Some of the kits were shipped at the incorrect shipping temperature. They were shipped ambient, but should have been shipped frozen.
Firm Action
Abbott telephoned the laboratory end users on 10/17/05, informing them that Abbott discovered that ARCHITECT STAT Myoglobin Calibrators, lot V96128, were shipped to their facilities using the incorrect shipping conditions between August 29th and September 20th 2005. The kits should have been shipped frozen, but were shipped ambient. The error has been corrected. All shipments made after September 21, 2005 have been shipped correctly. Any data generated to date using this kit of calibrators is acceptable, providing that a valid calibration was achieved and product labeling was followed. However, long term use of this kit may result in calibration failures or controls out of range. The accounts were requested to mark the kit(s) of this lot currently in inventory to distinguish them from the replacement kit(s) Abbott will be sending to them. As a precaution, Abbott requested the labs to discontinue use of these kits and destroy them the replacement kits are received.
Root Cause
Other
Distribution Pattern
Texas, Washington, Montana, Pennsylvania and Kentucky, and internationally to Korea
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.