Brilliance CT (computed tomography) scanners, Big Bore configuration, Model number 728243. software version 2.0 The products are intended to provide cross sectional images of the human body and visualization of the internal organs of the body.

FDA Recall Z-0400-2016 · Posted 2015-12-10

Recall StatusTerminated
Product CodeJAK
Date Initiated2006-04-25
Date Terminated2015-12-15
Product Quantity54 units

Reason for Recall

Brilliance Big Bore- An anomaly was identified with software version 2.0 tumor localization application when printing work sheets. this causes information regarding relative locations of marked regions of interest on the patient work sheets to be incorrect and may cause mistreatment.

Firm Action

Philips Medical Systems sent a product safety notice to customers and corrective software for the issue was released to the field on July 25, 2006. Service Reps will contact each customer to install the new version of software. For questions regarding this recall call 440-483-7600.

Root Cause

Software design

Distribution Pattern

Worldwide distribution, including US states of CA and NV.