Terminated

Simpurity HYDROGEL ABSORBENT SHEET WOUND DRESSING, 4" X 5", 10 Pads, Catalogue number SNS58315

FDA Recall Z-0400-2019 · Posted None

Recalling FirmSafe N Simple, LLC
Recall StatusTerminated
Product CodeNAE
Date Initiated2018-08-22
Date Terminated2020-03-17
Product Quantity104 cases

Reason for Recall

The sterilization process used was unqualified.

Firm Action

Safe n Simple extended the recall to the only customer that ordered the product in the lot numbers (see table). The firm provided a written notice to the customer mentioned in the table with specific instructions along with a phone conversation. Safe n Simple instructed the customer to provide the inventory level of SNS58315 and SNS58345 specific to lot numbers 18060564 and 18060563 respectively and to return such product to Safe N Simple LLC at Safe N Simple LLC expense. The effectiveness of the recall was good since the customer immediately responded by shipping all their inventory back to us.

Root Cause

Process control

Distribution Pattern

CA