BD Insyte Autoguard BC Shielded IV Catheter with Blood Control Technology 20GA X 1.00in, REF: 382533

FDA Recall Z-0400-2024 · Posted 2023-11-24

Recall StatusOpen, Classified
Product CodeFOZ
Date Initiated2023-10-04
Product Quantity528,000

Reason for Recall

Over-the-needle, intravascular catheters may have needles that do not retract properly into the barrel, which could result in contaminated needlestick injury, blood leaking from the insertion site or catheter, and repeat catheter insertion attempts, due to catheter being inadvertently removed from the insertion site.

Firm Action

On 10/4/23, recall notices were emailed to customers and distributors who were asked to do the following: 1) Destroy all unused product subject to the recall following your institution's process for destruction. 2) For indwelling catheters, no further action is needed. 3) Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 4) Complete and return the customer response form via email to [email protected] Customers with recall questions, product complaints, or technical questions can call 1-844-823-5433, say recall when prompted, Mon-Fri 8:00am to 5:00PM CT, or email [email protected] In addition, distributors are asked to identify all customers within their distribution network that purchased affected product. Then distributors should provide a copy of the customer letter to these customers and advise them of the recall.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the states of CA, NC, MI, WA, CO, FL, NJ, NM, IL, KY, TN, OH, TX, IN, LA, IA, MS, MO, NE, MD, NY, PA, GA, VA, ME, SC, RI, MA, PR.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.