STERRAD 100NX, Product Code: 10104 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

FDA Recall Z-0402-2014 · Posted 2013-11-26

Recall StatusTerminated
Product CodeMLR
Date Initiated2013-11-11
Date Terminated2014-01-22
Product QuantityTotal 4,381 units

Reason for Recall

Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.

Firm Action

Advanced Sterilization Products sent an Urgent Medical Device Field Safety Notificaiton letter on November 11, 2013, to all affected customers. The letter informed customers of the problems identified and the actions to be taken. Customers were instructed to provide the notice to anyone in their facility that needs to be informed and post this notice where appropriate. Customers were instructed to contact ASP at 1-888-783-7723 for any suspected problems. ASP is collaborating with KARL STORZ Endoskope to determine the sterilization impact of the suction port lumen material. Customers are instructed to contact KARL STORZ Endoskope Technical Support at (800)421-0837, ext 5350 for alternative reprocessing instructions. For questions regarding this recall call 949-453-6400.

Root Cause

Labeling design

Distribution Pattern

Nationwide Distribution